Water-equivalence of gel dosimeters for radiology medical imaging

Primer Autor
Valente, M.
Co-autores
Vedelago, J.#Chacon, D.#Mattea, F.#Velasquez, J.#Perez, P.
Título
Water-equivalence of gel dosimeters for radiology medical imaging
Editorial
PERGAMON-ELSEVIER SCIENCE LTD
Revista
APPLIED RADIATION AND ISOTOPES
Lenguaje
en
Resumen
International dosimetry protocols are based on determinations of absorbed dose to water. Ideally, the phantom material should be water equivalent, that is, it should have the same absorption and scatter properties as water. This study presents theoretical, experimental and Monte Carlo modeling of water-equivalence of Fricke and polymer (NIPAM, PAGAT and itaconic acid ITABIS) gel dosimeters. Mass and electronic densities along with effective atomic number were calculated by means of theoretical approaches. Samples were scanned by standard computed tomography. Photon mass attenuation coefficients and electron stopping powers were examined. Theoretical, Monte Carlo and experimental results confirmed good water-equivalence for all gel dosimeters. Overall variations with respect to water in the low energy radiology range (up to 130 kVp) were found to be less than 3% in average.
Tipo de Recurso
artículo original#presentación de congreso
Description
This study was partially financed by CONICET (Argentina) project ESPORA I - PIP 112-20130100658CO, SeCyT-UNC (Argentina) project DOSCOM I and DI-UFRO (Chile) projects DI-16-6008 and DI-17-1003. Authors are grateful to ICOS radiology technicians, especially Camila Vidal Bravo.
Este estudio fue financiado parcialmente por el proyecto ESPORA I - PIP 112-20130100658CO del CONICET (Argentina), el proyecto DOSCOM I de la SeCyT-UNC (Argentina) y los proyectos DI-16-6008 y DI-17-1003 de la DI-UFRO (Chile). Los autores agradecen a los técnicos de radiología de ICOS, especialmente a Camila Vidal Bravo.
doi
10.1016/j.apradiso.2018.03.005
Formato Recurso
pdf
Palabras Claves
Gel dosimetry# Water-equivalence# Computed tomography
Ubicación del archivo
http://dx.doi.org/10.1016/j.apradiso.2018.03.005
Categoría OCDE
Chemistry, Inorganic & Nuclear# Nuclear Science & Technology# Radiology, Nuclear Medicine & Medical Imaging
Materias
Dosimetría en gel# Equivalencia de agua# tomografía computarizada
Disciplinas de la OCDE
Radiología, Medicina Nuclear y de Imágenes
Otras Especialidades de la Física
Físico-Química
Id de Web of Science
WOS:000449448300031
Título de la cita (Recomendado-único)
Water-equivalence of gel dosimeters for radiology medical imaging
Identificador relacionado
http://hdl.handle.net/11336/91615
Identificador del recurso (Mandatado-único)
artículo original#presentación de congreso
Versión del recurso (Recomendado-único)
version publicada
Editorial
PERGAMON-ELSEVIER SCIENCE LTD
Revista/Libro
APPLIED RADIATION AND ISOTOPES
Categoría WOS
Química, Inorgánica y Nuclear# Ciencia y tecnología nucleares# Radiología, Medicina Nuclear e Imágenes Médicas
ISSN
0969-8043
Idioma
en
Referencia del Financiador (Mandatado si es aplicable-repetible)
CONICET ESPORA I - PIP 112-20130100658CO#UFRO DI-16-6008#UFRO DI-17-1003#ICOS
CONICET (Argentina) ESPORA I - PIP 112-20130100658CO]
SeCyT-UNC (Argentina) DOSCOM I
UFRO DI-16-6008
UFRO DI-17-1003
Descripción
This study was partially financed by CONICET (Argentina) project ESPORA I - PIP 112-20130100658CO, SeCyT-UNC (Argentina) project DOSCOM I and DI-UFRO (Chile) projects DI-16-6008 and DI-17-1003. Authors are grateful to ICOS radiology technicians, especially Camila Vidal Bravo.
Formato
pdf
Tipo de ruta
hibrida#verde
Access Rights
metadata
Derechos de acceso
metadata
Página de inicio (Recomendado-único)
34
Página final (Recomendado-único)
50
Revisa las metricas alternativas de Almetrics
Revisa las citaciones de Dimensions