Evaluation of Failed Implants by Metallographic and Energy Dispersive X-ray Analysis

Primer Autor
Soares de Alcantara Pinto, Clarice Maia
Co-autores
Goulart, Douglas Rangel#Asprino, Luciana#Olate, Sergio#de Moraes, Marcio
Título
Evaluation of Failed Implants by Metallographic and Energy Dispersive X-ray Analysis
Editorial
LIPPINCOTT WILLIAMS & WILKINS
Revista
IMPLANT DENTISTRY
Lenguaje
en
Resumen
"Objective: To evaluate the microscopic structure and chemical composition of titanium dental implants removed from patients with implant failure. Materials and Methods: Thirtyeight dental implants removed from 29 patients between January 2012 and September 2014 were studied by metallographic analysis, and 7 of these implants were also studied by energy-dispersive x-ray (EDX) analysis. Medical records of these patients were examined. Results: Dental implants were removed in a median of 54.6 +/- 88.1 weeks after insertion. The lack of osseointegration without symptomatology or signs of infection was the main reason for implant failure (55.17%). Seventeen patients pre-sented implant failure in the maxilla and 12 patients in the mandible. Metallographic analysis revealed that all the dental implants were manufactured from commercially pure titanium, presenting microscopic uniform appearance, with no ""alpha case"". The implants studied by EDX analysis presented between 99.85% and 99.87% of titanium and 0.13% and 0.15% of iron by mass. All the implants evaluated were within the ASTM specifications in both analyses. Conclusion: There was no relationship between dental implant failure and the microscopic structure and chemical composition of these devices."
Tipo de Recurso
Artículo original
doi
10.1097/ID.0000000000000700
Formato Recurso
pdf
Palabras Claves
dental materialsosseointegration# prostheses and implants# titanium
Ubicación del archivo
http://dx.doi.org/10.1097/ID.0000000000000700
Categoría OCDE
Dentistry, Oral Surgery & Medicine
Materias
osteointegración de materiales dentales# prótesis e implantes# titanio
Disciplinas de la OCDE
Odontología, Cirugía y Medicina Oral
Ingeniería de los Materiales
Química Analítica
Id de Web of Science
WOS:000429311200012
Título de la cita (Recomendado-único)
Evaluation of Failed Implants by Metallographic and Energy Dispersive X-ray Analysis
Identificador del recurso (Mandatado-único)
Artículo original
Versión del recurso (Recomendado-único)
version publicada
Editorial
LIPPINCOTT WILLIAMS & WILKINS
Revista/Libro
IMPLANT DENTISTRY
Categoría WOS
Odontología, Cirugía Bucal y Medicina
ISSN
1056-6163
Idioma
en
Formato
pdf
Tipo de ruta
hibrida
Access Rights
acceso abierto
Derechos de acceso
acceso abierto
Página de inicio (Recomendado-único)
305
Página final (Recomendado-único)
313
Revisa las metricas alternativas de Almetrics
Revisa las citaciones de Dimensions